ASTM F2063
Nickel-Titanium Shape Memory Alloys for Medical Devices and Surgical Implants
ASTM F2063 is a standard specification for Wrought Nickel-Titanium Shape Memory Alloys for Medical Devices and Surgical Implants. It covers the chemical, physical, mechanical, and metallurgical requirements for wrought nickel-titanium bar, flat rolled products, and tubes containing nominally 54.5- to 57.0-weight percent nickel and used for the manufacture of medical devices and surgical implants.
Chemical Composition
ASTM F2063-12 Chemical Requirements:
|
Element |
Composition,% (mass/mass) |
|
Nickel |
54.5 to 57.0 |
|
Carbon (Maximum) |
0.050 |
|
Cobalt (Maximum) |
0.050 |
|
Copper (Maximum) |
0.010 |
|
Chromium(Maximum) |
0.010 |
|
Hydrogen(Maximum) |
0.005 |
|
Iron (Maximum) |
0.050 |
|
Niobium (Maximum) |
0.025 |
|
Nitrogen plus Oxygen(Maximum) |
0.050 |
|
Titanium |
Balance |
Analysis Tolerances
|
Element |
Tolerance Under the Minimum Limit or Over the Maximum Limit,%(mass/mass) |
|
Carbon |
0.002 |
|
Cobalt |
0.001 |
|
Copper |
0.001 |
|
Chromium |
0.001 |
|
Hydrogen |
0.0005 |
|
Iron |
0.01 |
|
Nickel |
0.2 under min; 0.2 over max |
|
Niobium |
0.004 |
|
Nitrogen |
0.004 |
|
Oxygen |
0.004 |
Tolerance Requirements
ASTM F2063-12
|
As (°C) |
Tolerance Range (°C) |
|
≥70 |
± 7 |
|
50 < As < 70 |
± 10 |
|
≤–50 |
± 15 |
Annealed Mechanical Properties
ASTM F2063-12
|
Diameter mm(in) |
Tensile strength MPa, Minimum |
Elongation in 50.0 mm [1.97 in.] or 4 D, % Minimum |
|
≤50 (<1.97 in) |
551 (79.9 KSI) |
15 |
|
>50(>1.97 in) |
551 (79.9 KSI) |
10 |
Copyright © 2024 Changzhou Bokang Special Material Technology Co., Ltd. All Rights Reserved.